Beer–Lambert Law Calculator

Pick the unknown — absorbance, molar absorption coefficient, path length or concentration — and enter the other three in the units you have. The calculator solves A = εlc, shows the working, and gives the matching percent transmittance. A second mode converts between absorbance and transmittance on its own. Free, no sign-up.

What do you want to calculate?
Solve for
Try an example

ε at the wavelength you measure. M⁻¹ cm⁻¹ is the same as L mol⁻¹ cm⁻¹; the SI unit m² mol⁻¹ is 10 times smaller in number.

Results

0.622

Absorbance (A)

23.9%

Transmittance (%T)

76.1%

Light absorbed

The working

A = ε × l × c

Concentration (c): 100. µM = 1.0000 × 10⁻⁴ mol/L

A = 6220 × 1 × 1.0000 × 10⁻⁴ = 0.6220

%T = 10^(2 − A) = 10^(2 − 0.622) = 23.878113%

How to Use the Beer–Lambert Law Calculator

  1. 1. Choose the mode. "A = εlc for any term" solves the Beer–Lambert law; "Absorbance ↔ transmittance" converts a single reading.
  2. 2. Pick what you want to find. Most often it is the concentration of a sample from its measured absorbance, but ε and l can be found too. The answer box turns into a unit menu.
  3. 3. Enter the rest with units. The molar absorption coefficient can be in M⁻¹ cm⁻¹, mM⁻¹ cm⁻¹ or m² mol⁻¹; path length in cm, mm or m; concentration from nM to M. A reading given as %T or as a fraction T is converted to absorbance first.
  4. 4. Check the range note. The calculator warns when absorbance is below about 0.1 or above about 1, where most instruments lose accuracy.

The Beer–Lambert Law

Law

A = ε × l × c

Absorbance

A = log₁₀(I₀ / I) = −log₁₀ T

Transmittance

%T = 100 × 10^(−A)

Concentration

c = A / (ε × l)

A is the absorbance (a pure number), ε the molar absorption coefficient at the wavelength used, l the path length through the sample (1 cm for a standard cuvette) and c the concentration of the absorbing species. IUPAC calls ε the molar (decadic) absorption coefficient; "molar extinction coefficient" is the older, still common name, and IUPAC advises against "molar absorptivity". Its usual unit is L mol⁻¹ cm⁻¹, which is the same as M⁻¹ cm⁻¹; the SI unit is m² mol⁻¹, and 1 m² mol⁻¹ equals 10 M⁻¹ cm⁻¹.

Absorbance is logarithmic: each unit of A removes 90% of what is left. That is why a reading of 2 means only 1% of the light gets through, and why absorbance adds up for stacked samples while transmittance multiplies.

Absorbance (A)Transmittance (%T)Light absorbed
01000%
0.179.420.6%
0.350.149.9%
0.531.668.4%
11090%
2199%
30.199.9%

Worked Examples

Absorbance of NADH (the default inputs). NADH absorbs at 340 nm with ε ≈ 6220 M⁻¹ cm⁻¹, a value widely used in enzyme assays (some references give about 6300, so use your lab's). A 100. µM solution in a 1.00 cm cuvette has A = 6220 × 1.00 × 1.00 × 10⁻⁴ = 0.622, which is 23.9% transmittance.

Concentration from a reading. A sample of the same compound reads A = 0.450. Then c = 0.450 ÷ (6220 × 1) = 7.23 × 10⁻⁵ M, or 72.3 µM. Taking the reading as %T instead: 35.5% transmittance is the same A = 0.450.

Finding ε. A 40.0 µM solution gives A = 0.800 in a 1.00 cm cell, so ε = 0.800 ÷ (1.00 × 4.00 × 10⁻⁵) = 20 000 M⁻¹ cm⁻¹. Halving the path length to 0.5 cm would halve the absorbance.

When the Law Stops Working

  • High absorbance. Above A ≈ 1 the detector sees mostly stray light, and the plot of A against c bends over. Dilute and remeasure.
  • Concentrated solutions. Above about 10 mM, molecules interact and ε changes.
  • Wide bandwidth. The law assumes narrow, monochromatic light; IUPAC states it holds only when the bandwidth is narrow compared with the spectral lines.
  • Chemistry in the cuvette. Dimerization, ionization or a pH-sensitive dye changes the absorbing species as you dilute it.
  • Scattering and cloudy samples. Light scattered out of the beam is counted as absorbed. Filter or centrifuge first.

If you have several standards rather than a literature ε, fit a calibration line instead: the Standard Curve Calculator takes absorbances of known concentrations and reads off the unknown. Remember to subtract a blank, and keep unknowns inside the range of the standards.

Frequently Asked Questions

What is the Beer–Lambert law formula?
A = εlc: absorbance equals the molar absorption coefficient times the path length times the concentration. It says that a dilute solution absorbs light in proportion to how much absorbing material the beam passes through.
How do I calculate concentration from absorbance?
Divide the absorbance by ε and by the path length: c = A ÷ (ε × l). With A = 0.450, ε = 6220 M⁻¹ cm⁻¹ and l = 1 cm, c = 7.23 × 10⁻⁵ M. Make sure the absorbance has been blank-corrected.
How do I convert absorbance to transmittance?
Transmittance is 10 raised to the power of minus the absorbance, T = 10^(−A); multiply by 100 for percent. So A = 1 is 10% T and A = 2 is 1% T. In reverse, A = 2 − log₁₀(%T).
What is a good absorbance range for a spectrophotometer?
Roughly 0.1 to 1.0, equal to about 80% down to 10% transmittance. Below 0.1 the signal is close to noise, and above about 1 stray light makes the response non-linear. Dilute or change the path length to bring a sample into range.
What is the difference between the molar extinction coefficient and the molar absorption coefficient?
They are the same quantity. IUPAC recommends "molar (decadic) absorption coefficient"; "molar extinction coefficient" is the older name. "Molar absorptivity" is also used, but IUPAC advises against it.
Why is my calibration curve not a straight line at high concentration?
The Beer–Lambert law is a limiting law. At high absorbance stray light, wide bandwidth, molecular interactions or chemical equilibria bend the curve. Keep standards below about A = 1, or dilute the unknown into the linear region.